      Characterization of textile electrodes and conductors using standardized measurement setups.

      Physiological measurement
      Algorithms, Electric Impedance, Electricity, Electrodes, Electronics, Medical, instrumentation, Humans, Mechanical Phenomena, Models, Biological, Photoperiod, Reproducibility of Results, Skin Physiological Phenomena, Textiles, Time Factors

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Textile electrodes and conductors are being developed and used in different monitoring scenarios, such as ECG or bioimpedance spectroscopy measurements. Compared to standard materials, conductive textile materials offer improved wearing comfort and enable long-term measurements. Unfortunately, the development and investigation of such materials often suffers from the non-reproducibility of the test scenarios. For example, the materials are generally tested on human skin which is difficult since the properties of human skin differ for each person and can change within hours. This study presents two test setups which offer reproducible measurement procedures for the systematic analysis of textile electrodes and conductors. The electrode test setup was designed with a special skin dummy which allows investigation of not only the electrical properties of textile electrodes but also the contact behavior between electrode and skin. Using both test setups, eight textile electrodes and five textile conductors were analyzed and compared.

          Related collections

          Author and article information

          Comments

          Comment on this article