      Comparison of the accuracy of the Binkhorst, Colenbrander, and SRK implant power prediction formulas.

      Read this article at

      ScienceOpenPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The prediction accuracy of the Binkhorst, Colenbrander, and SRK formulas were compared in five series from different sources totalling 654 cases. The SRK formula was superior to both the other formulas by having a smaller average error per case, a smaller range of error from highest minus to highest plus, and a smaller proportion of cases with greater than 2 diopters (D) of error in all five series studied. In four of the five series, the SRK formula also had the greatest proportion of cases with less than 1 D of error. The Colenbrander formula was superior to the Binkhorst formula in all five series with regard to average error, proportion of cases with less than 1 D of error, and proportion of cases with more than 2 D of error. The range of error from highest minus to highest plus was equivalent with the two formulas. All available published literature on the accuracy of implant power prediction formulas was reviewed and appears to support our findings.

          Related collections

          Author and article information

          Journal
          J Am Intraocul Implant Soc
          Journal - American Intra-Ocular Implant Society
          0146-2776
          0146-2776
          1981
          : 7
          : 4
          Article
          7349649
          3d1eddb5-425f-4265-bdae-d74b8df73f01
          History

          Comments

          Comment on this article