      Comparative cost-efficiency of the EVOTECH endoscope cleaner and reprocessor versus manual cleaning plus automated endoscope reprocessing in a real-world Canadian hospital endoscopy setting

      research-article
      1 , , 2
      BMC Gastroenterology
      BioMed Central

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Reprocessing of endoscopes generally requires labour-intensive manual cleaning followed by high-level disinfection in an automated endoscope reprocessor (AER). EVOTECH Endoscope Cleaner and Reprocessor (ECR) is approved for fully automated cleaning and disinfection whereas AERs require manual cleaning prior to the high-level disinfection procedure. The purpose of this economic evaluation was to determine the cost-efficiency of the ECR versus AER methods of endoscopy reprocessing in an actual practice setting.

          Methods

          A time and motion study was conducted at a Canadian hospital to collect data on the personnel resources and consumable supplies costs associated with the use of EVOTECH ECR versus manual cleaning followed by AER with Medivators DSD-201. Reprocessing of all endoscopes was observed and timed for both reprocessor types over three days. Laboratory staff members were interviewed regarding the consumption and cost of all disposable supplies and equipment. Exact Wilcoxon rank sum test was used for assessing differences in total cycle reprocessing time.

          Results

          Endoscope reprocessing was significantly shorter with the ECR than with manual cleaning followed by AER. The differences in median time were 12.46 minutes per colonoscope (p < 0.0001), 6.31 minutes per gastroscope (p < 0.0001), and 5.66 minutes per bronchoscope (p = 0.0040). Almost 2 hours of direct labour time was saved daily with the ECR. The total per cycle cost of consumables and labour for maintenance was slightly higher for EVOTECH ECR versus manual cleaning followed by AER ($8.91 versus $8.31, respectively). Including the cost of direct labour time consumed in reprocessing scopes, the per cycle and annual costs of using the EVOTECH ECR was less than the cost of manual cleaning followed by AER disinfection ($11.50 versus $11.88).

          Conclusions

          The EVOTECH ECR was more efficient and less costly to use for the reprocessing of endoscopes than manual cleaning followed by AER disinfection. Although the cost of consumable supplies required to reprocess endoscopes with EVOTECH ECR was slightly higher, the value of the labour time saved with EVOTECH ECR more than offset the additional consumables cost. The increased efficiency with EVOTECH ECR could lead to even further cost-savings by shifting endoscopy laboratory personnel responsibilities but further study is required.

          Related collections

          Most cited references10

          • Record: found
          • Abstract: not found
          • Article: not found

          Interventional pulmonary procedures: Guidelines from the American College of Chest Physicians.

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Transmission of infection by gastrointestinal endoscopy and bronchoscopy.

            To review reports on the transmission of infections by flexible gastrointestinal endoscopy and bronchoscopy in order to determine common infecting microorganisms, circumstances of transmission, and methods of risk reduction. Relevant English-language articles were identified through prominent review articles and a MEDLINE search (1966 to July 1992); additional references were selected from the bibliographies of identified articles. All selected articles related to transmission of infection by gastrointestinal endoscopy or bronchoscopy; 265 articles were reviewed in detail. Two hundred and eighty-one infections were transmitted by gastrointestinal endoscopy, and 96 were transmitted by gastrointestinal endoscopy, spectrum of these infections ranged from asymptomatic colonization to death. Salmonella species and Pseudomonas aeruginosa were repeatedly identified as the causative agents of infections transmitted by gastrointestinal endoscopy, and Mycobacterium tuberculosis, atypical mycobacteria, and P. aeruginosa were the most common causes of infections transmitted by bronchoscopy. One case of hepatitis B virus transmission via gastrointestinal endoscopy was documented. Major reasons for transmission were improper cleaning and disinfection procedures; the contamination of endoscopes by automatic washers; and an inability to decontaminate endoscopes, despite the use of standard disinfection techniques, because of their complex channel and valve systems. The most common agents of infection transmitted by endoscopy are Salmonella, Pseudomonas, and Mycobacterium species. To prevent endoscopic transmission of infections, recommended disinfection guidelines must be followed, the effectiveness of automatic washers must be carefully monitored, and improvements in endoscope design are needed to facilitate effective cleaning and disinfection.
              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              Patient-to-patient transmission of hepatitis C virus during colonoscopy.

                Bookmark

                Author and article information

                Journal
                BMC Gastroenterol
                BMC Gastroenterology
                BioMed Central
                1471-230X
                2011
                3 October 2011
                : 11
                : 105
                Affiliations
                [1 ]VALORE Research, Toronto, Ontario, Canada
                [2 ]Johnson and Johnson Medical Products, Department of Health Economics & Outcomes Research, Markham, Ontario, Canada
                Article
                1471-230X-11-105
                10.1186/1471-230X-11-105
                3198958
                21967345
                71a25e33-bbcd-4c65-b80c-aa81ca21f32a
                Copyright ©2011 Forte and Shum; licensee BioMed Central Ltd.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 7 December 2010
                : 3 October 2011
                Categories
                Research Article

                Gastroenterology & Hepatology
                Gastroenterology & Hepatology

                Comments

                Comment on this article