13
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      The pacemaker-twiddler’s syndrome: an infrequent cause of pacemaker failure

      case-report

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          The pacemaker-twiddler’s syndrome is an uncommon cause of pacemaker malfunction. It occurs due to unintentional or deliberate manipulation of the pacemaker pulse generator within its skin pocket by the patient. This causes coiling of the lead and its dislodgement, resulting in failure of ventricular pacing. More commonly reported among elderly females with impaired cognition, the phenomenon usually occurs in the first year following pacemaker implantation. Treatment involves repositioning of the dislodged leads and suture fixation of the lead and pulse generator within its pocket.

          Case presentation

          An 87 year old Bangladeshi lady who underwent a single chamber ventricular pacemaker (VVI mode: i.e. ventricle paced, ventricle sensed, inhibitory mode) implantation with the indication of complete heart block, and presented to us again 7 weeks later, with syncopal attacks. She admitted to repeatedly manipulating the pacemaker generator in her left pectoral region. Physical examination revealed a heart rate of 42 beats/minute, blood pressure 140/80 mmHg and bilateral crackles on lung auscultation. She had no cognitive deficit.

          An immediate electrocardiogram showed complete heart block with pacemaker spikes and failure to capture. Chest X-ray showed coiled and retracted right ventricular lead and rotated pulse generator. An emergent temporary pace maker was set at a rate of 60 beats per minute. Subsequently, she underwent successful lead repositioning with strong counselling to avoid further twiddling.

          Conclusion

          Twiddler’s syndrome should be considered as a cause of pacemaker failure in elderly patients presenting with bradyarrythmias following pacemaker implantation. Chest X-ray and electrocardiograms are simple and easily-available first line investigations for its diagnosis. Lead repositioning is required, however proper patient education and counselling against further manipulation is paramount to long-term management.

          Related collections

          Most cited references10

          • Record: found
          • Abstract: not found
          • Article: not found

          The pacemaker-twiddler's syndrome: a new complication of implantable transvenous pacemakers.

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Complications of permanent transvenous cardiac pacing: a 14-year review of all transvenous pacemakers inserted at one community hospital.

            Five hundred and eighty-nine consecutive transvenous pacemakers implanted between January 1, 1971 and January 1, 1985 were reviewed. A complication incidence of 17% early in the study period (1971-1977) was caused by electrode displacement and perforation. As of 1975, smaller-sized electrodes with endocardial fixation were used and were associated with a significant decrease in this complication incidence to 2% by 1985. All other complications occurring during the entire study period (1971-1985) occurred at a frequency of 3% (20/589). Infection occurred in 3 of 589 cases; there were no extrusions or erosions. No mortality was attributed to pacemaker insertion. The "twiddler" syndrome seems to be increasing in frequency. Transvenous permanent pacing can be accomplished today with a low complication rate of 5% or less. The majority of these complications are minor and can be corrected easily. Our studies suggest that the reduced complication incidence is mainly related to improved technology.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Early pacemaker twiddler syndrome.

              Twiddler's syndrome is a well-known complication of pacemaker treatment. It was first described by Bayliss et al. when a patient manipulated and rotated the pulse generator in the pocket so many turns that it resulted in lead dislodgment, diaphragmatic stimulation and loss of capture. In this case report we present a patient who managed to rotate her dual chamber pulse generator so quickly after implantation that exit block occurred within 17 h. She had wound the two leads as far as their tips in a perfect formation around the pulse generator.
                Bookmark

                Author and article information

                Contributors
                0088 1711859385 , riponsala@gmail.com
                aaysha.cader@gmail.com
                nasrin_jhumur@hotmail.com
                mashhudzia@yahoo.com
                Journal
                BMC Res Notes
                BMC Res Notes
                BMC Research Notes
                BioMed Central (London )
                1756-0500
                20 January 2016
                20 January 2016
                2016
                : 9
                : 32
                Affiliations
                Department of Cardiology, Ibrahim Cardiac Hospital and Research Institute (ICHRI), Dhaka, Bangladesh
                Article
                1818
                10.1186/s13104-015-1818-0
                4721019
                26790626
                7e716989-3e64-4273-8f59-d2f2f5d2573c
                © Salahuddin et al. 2016

                Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License ( http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 28 May 2015
                : 17 December 2015
                Categories
                Case Report
                Custom metadata
                © The Author(s) 2016

                Medicine
                twiddler’s syndrome,pacemaker failure,lead dislodgement
                Medicine
                twiddler’s syndrome, pacemaker failure, lead dislodgement

                Comments

                Comment on this article