      Is noncoplanar plan more robust to inter‐fractional variations than coplanar plan in treating bilateral HN tumors with pencil‐beam scanning proton beams?

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          Noncoplanar plans (NCPs) are commonly used for proton treatment of bilateral head and neck (HN) malignancies. NCP requires additional verification setup imaging between beams to correct residual errors of robotic couch motion, which increases imaging dose and total treatment time. This study compared the quality and robustness of NCPs with those of coplanar plans (CPs).

          Methods and Materials

          Under an IRB‐approved study, CPs were created retrospectively for 10 bilateral HN patients previously treated with NCPs maintaining identical beam geometry of the original plan but excluding couch rotations. Plan robustness to the inter‐fractional variation (IV) of both plans was evaluated through the Dose Volume Histograms (DVH) of weekly quality assurance CT (QACT) sets (39 total). In addition, delivery efficiency for both plans was compared using total treatment time (TTT) and beam‐on time (BOT).

          Results

          No significant differences in plan quality were observed in terms of clinical target volume (CTV) coverage (D95) or organ‐at‐risk (OAR) doses ( p > 0.4 for all CTVs and OARs). No significant advantage of NCPs in the robustness to IV was found over CP, either. Changes in D95 of QA plans showed a linear correlation (slope = 1.006, R 2 > 0.99) between NCP and CP for three CTV data points (CTV1, CTV2, and CTV3) in each QA plan (117 data points for 39 QA plans). NCPs showed significantly higher beam delivery time than CPs for TTT (539 ± 50 vs. 897 ± 142 s; p < 0.001); however, no significant differences were observed for BOT.

          Conclusion

          NCPs are not more robust to IV than CPs when treating bilateral HN tumors with pencil‐beam scanning proton beams. CPs showed plan quality and robustness similar to NCPs while reduced treatment time (∼6 min). This suggests that CPs may be a more efficient planning technique for bilateral HN cancer proton therapy.

          Related collections

          Most cited references19

          • Record: found
          • Abstract: found
          • Article: not found

          Intensity modulated proton therapy (IMPT) – The future of IMRT for head and neck cancer

          Radiation therapy plays an integral role in the management of head and neck cancers (HNCs). While most HNC patients have historically been treated with photon-based radiation techniques such as intensity modulated radiation therapy (IMRT), there is a growing awareness of the potential clinical benefit of proton therapy over IMRT in the definitive, postoperative and reirradiation settings given the unique physical properties of protons. Intensity modulated proton therapy (IMPT), also known as “pencil beam proton therapy,” is a sophisticated mode of proton therapy that is analogous to IMRT and an active area of investigation in cancer care. Multifield optimization IMPT allows for high quality plans that can target superficially located HNCs as well as large neck volumes while significantly reducing integral doses. Several dosimetric studies have demonstrated the superiority of IMPT over IMRT to improve dose sparing of nearby organs such as the larynx, salivary glands, and esophagus. Evidence of the clinical translation of these dosimetric advantages has been demonstrated with documented toxicity reductions (such as decreased feeding tube dependency) after IMPT for patients with HNCs. While there are relative challenges to IMPT planning that exist today such as particle range uncertainties and high sensitivity to anatomical changes, ongoing investigations in image-guidance techniques and robust optimization methods are promising. A systematic approach towards utilizing IMPT and additional prospective studies are also necessary in order to more accurately estimate the clinical benefit of IMPT over IMRT and passive proton therapy on a case-by-case basis for patients with sub-site specific HNCs.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Proton therapy for head and neck cancer: expanding the therapeutic window.

            Use of proton beam therapy has expanded, with the number of proton centres rapidly increasing not only in the USA but also worldwide. The physical characteristics of the proton beam offer important advantages versus widely used photon techniques in terms of radiation precision. In head and neck cancer in particular, proton beam therapy is uniquely suited for the complex anatomy of tumours and sensitive surrounding organs. De-intensification and personalisation of treatment to limit toxicity are of renewed importance in the context of human papilloma virus-associated disease, in which young patients will be cured but bear the consequences of adverse effects for decades. Comparisons of radiation dose distributions between photon and proton techniques suggest considerable benefit in terms of toxicity sparing, but this has only recently been confirmed by substantial clinical data. In this Review, we attempt to define the role of this method in the contemporary multidisciplinary management of various types of head and neck cancer.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              Robust Intensity Modulated Proton Therapy (IMPT) Increases Estimated Clinical Benefit in Head and Neck Cancer Patients

              Purpose To compare the clinical benefit of robust optimized Intensity Modulated Proton Therapy (minimax IMPT) with current photon Intensity Modulated Radiation Therapy (IMRT) and PTV-based IMPT for head and neck cancer (HNC) patients. The clinical benefit is quantified in terms of both Normal Tissue Complication Probability (NTCP) and target coverage in the case of setup and range errors. Methods and Materials For 10 HNC patients, PTV-based IMRT (7 fields), minimax and PTV-based IMPT (2, 3, 4, 5 and 7 fields) plans were tested on robustness. Robust optimized plans differed from PTV-based plans in that they target the CTV and penalize possible error scenarios, instead of using the static isotropic CTV-PTV margin. Perturbed dose distributions of all plans were acquired by simulating in total 8060 setup (±3.5 mm) and range error (±3%) combinations. NTCP models for xerostomia and dysphagia were used to predict the clinical benefit of IMPT versus IMRT. Results The robustness criterion was met in the IMRT and minimax IMPT plans in all error scenarios, but this was only the case in 1 of 40 PTV-based IMPT plans. Seven (out of 10) patients had relatively large NTCP reductions in minimax IMPT plans compared to IMRT. For these patients, xerostomia and dysphagia NTCP values were reduced by 17.0% (95% CI; 13.0–21.1) and 8.1% (95% CI; 4.9–11.2) on average with minimax IMPT. Increasing the number of fields did not contribute to plan robustness, but improved organ sparing. Conclusions The estimated clinical benefit in terms of NTCP of robust optimized (minimax) IMPT is greater than that of IMRT and PTV-based IMPT in HNC patients. Furthermore, the target coverage of minimax IMPT plans in the presence of errors was comparable to IMRT plans.
                Bookmark

                Author and article information

                Contributors
                byi@umm.edu
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                16 November 2023
                February 2024
                : 25
                : 2 ( doiID: 10.1002/acm2.v25.2 )
                : e14186
                Affiliations
                [ 1 ] Department of Radiation Oncology University of Maryland School of Medicine Baltimore Maryland USA
                [ 2 ] Maryland Proton Treatment Center Baltimore Maryland USA
                Author notes
                [*] [* ] Correspondence

                ByongYong Yi, Department of Radiation Oncology, University of Maryland School of Medicine, 22 S Greene St, Baltimore, MD 21201, USA.

                Email: byi@ 123456umm.edu

                Author information
                https://orcid.org/0000-0002-8805-0250
                https://orcid.org/0000-0002-6548-8321
                https://orcid.org/0000-0002-9527-0921
                https://orcid.org/0000-0002-8221-1631
                https://orcid.org/0000-0003-4333-648X
                https://orcid.org/0000-0003-1605-070X
                https://orcid.org/0000-0002-1295-0404
                https://orcid.org/0000-0003-0402-4707
                https://orcid.org/0000-0001-9007-3112
                Article
                ACM214186
                10.1002/acm2.14186
                10860533
                37974385
                7ef54bbe-fdfc-45b2-bb33-a0e0beec0ba4
                © 2023 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.

                This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 05 October 2023
                : 30 June 2023
                : 10 October 2023
                Page count
                Figures: 5, Tables: 3, Pages: 9, Words: 4359
                Categories
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                February 2024
                Converter:WILEY_ML3GV2_TO_JATSPMC version:6.3.6 mode:remove_FC converted:12.02.2024

                head & neck,noncoplanar plan,proton pbs
                head & neck, noncoplanar plan, proton pbs

                Comments

                Comment on this article