      Is a weekly qualitative picket fence test sufficient? A proposed alternate EPID‐based weekly MLC QA program

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          Well‐designed routine multileaf collimator (MLC) quality assurance (QA) is important to assure external‐beam radiation treatment delivery accuracy. This study evaluates the clinical necessity of a comprehensive weekly (C‐Weekly) MLC QA program compared to the American Association of Physics in Medicinerecommended weekly picket fence test (PF‐Weekly), based on our seven‐year experience with weekly MLC QA.

          Methods

          The C‐Weekly MLC QA program used in this study includes 5 tests to analyze: (1) absolute MLC leaf position; (2) interdigitation MLC leaf position; (3) picket fence MLC leaf positions at static gantry angle; (4) minimum leaf‐gap setting; and (5) volumetric‐modulated arc therapy delivery. A total of 20,226 QA images from 16,855 tests (3,371 tests × 5) for 11 linacs at 5 photon clinical sites from May 2014 to June 2021 were analyzed. Failure mode and effects analysis was performed with 5 failure modes related to the 5 tests. For each failure mode, a risk probability number (RPN) was calculated for a C‐Weekly and a PF‐Weekly MLC QA program. The probability of occurrence was evaluated from statistical analyses of the C‐Weekly MLC QA.

          Results

          The total number of failures for these 16,855 tests was 143 (0.9%): 39 (27.3%) for absolute MLC leaf position, 13 (9.1%) for interdigitation position, 9 (6.3%) for static gantry picket fence, 2 (1.4%) for minimum leaf‐gap setting, and 80 (55.9%) for VMAT delivery. RPN scores for PF‐Weekly MLC QA ranged from 60 to 192 and from 48 to 96 for C‐Weekly MLC QA.

          Conclusion

          RPNs for the 5 failure modes of MLC QA tests were quantitatively determined and analyzed. A comprehensive weekly MLC QA is imperative to lower the RPNs of the 5 failure modes to the desired level (<125); those from the PF‐Weekly MLC QA program were found to be higher (>125). This supports the clinical necessity for comprehensive weekly MLC QA.

          Related collections

          Most cited references27

          • Record: found
          • Abstract: found
          • Article: not found

          Task Group 142 report: quality assurance of medical accelerators.

          The task group (TG) for quality assurance of medical accelerators was constituted by the American Association of Physicists in Medicine's Science Council under the direction of the Radiation Therapy Committee and the Quality Assurance and Outcome Improvement Subcommittee. The task group (TG-142) had two main charges. First to update, as needed, recommendations of Table II of the AAPM TG-40 report on quality assurance and second, to add recommendations for asymmetric jaws, multileaf collimation (MLC), and dynamic/virtual wedges. The TG accomplished the update to TG-40, specifying new test and tolerances, and has added recommendations for not only the new ancillary delivery technologies but also for imaging devices that are part of the linear accelerator. The imaging devices include x-ray imaging, photon portal imaging, and cone-beam CT. The TG report was designed to account for the types of treatments delivered with the particular machine. For example, machines that are used for radiosurgery treatments or intensity-modulated radiotherapy (IMRT) require different tests and/or tolerances. There are specific recommendations for MLC quality assurance for machines performing IMRT. The report also gives recommendations as to action levels for the physicists to implement particular actions, whether they are inspection, scheduled action, or immediate and corrective action. The report is geared to be flexible for the physicist to customize the QA program depending on clinical utility. There are specific tables according to daily, monthly, and annual reviews, along with unique tables for wedge systems, MLC, and imaging checks. The report also gives specific recommendations regarding setup of a QA program by the physicist in regards to building a QA team, establishing procedures, training of personnel, documentation, and end-to-end system checks. The tabulated items of this report have been considerably expanded as compared with the original TG-40 report and the recommended tolerances accommodate differences in the intended use of the machine functionality (non-IMRT, IMRT, and stereotactic delivery).
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119.

            AAPM Task Group 119 has produced quantitative confidence limits as baseline expectation values for IMRT commissioning. A set of test cases was developed to assess the overall accuracy of planning and delivery of IMRT treatments. Each test uses contours of targets and avoidance structures drawn within rectangular phantoms. These tests were planned, delivered, measured, and analyzed by nine facilities using a variety of IMRT planning and delivery systems. Each facility had passed the Radiological Physics Center credentialing tests for IMRT. The agreement between the planned and measured doses was determined using ion chamber dosimetry in high and low dose regions, film dosimetry on coronal planes in the phantom with all fields delivered, and planar dosimetry for each field measured perpendicular to the central axis. The planar dose distributions were assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop confidence limits for the test results using the concept confidence limit = /mean/ + 1.96sigma. Other facilities can use the test protocol and results as a basis for comparison to this group. Locally derived confidence limits that substantially exceed these baseline values may indicate the need for improved IMRT commissioning.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218.

              Patient-specific IMRT QA measurements are important components of processes designed to identify discrepancies between calculated and delivered radiation doses. Discrepancy tolerance limits are neither well defined nor consistently applied across centers. The AAPM TG-218 report provides a comprehensive review aimed at improving the understanding and consistency of these processes as well as recommendations for methodologies and tolerance limits in patient-specific IMRT QA.
                Bookmark

                Author and article information

                Contributors
                byi@umm.edu
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                20 July 2022
                August 2022
                : 23
                : 8 ( doiID: 10.1002/acm2.v23.8 )
                : e13699
                Affiliations
                [ 1 ] Department of Radiation Oncology University of Maryland School of Medicine Baltimore Maryland USA
                Author notes
                [*] [* ] Correspondence

                ByongYong Yi, Department of Radiation Oncology, University of Maryland School of Medicine, 20 South Greene Street, Baltimore, MD 21201, USA.

                Email: byi@ 123456umm.edu

                Article
                ACM213699
                10.1002/acm2.13699
                9359020
                35856943
                82c1cb9b-e4ee-429a-8083-ada3b5ce3238
                © 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.

                This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 18 May 2022
                : 07 February 2021
                : 30 May 2022
                Page count
                Figures: 5, Tables: 4, Pages: 10, Words: 5417
                Categories
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                August 2022
                Converter:WILEY_ML3GV2_TO_JATSPMC version:6.1.7 mode:remove_FC converted:08.08.2022

                american association of physics in medicine task group 142,electronic portal imaging device,failure mode and effects analysis,leaf‐end leakage,multileaf collimator quality assurance,picket fence,volumetric‐modulated arc therapy

                Comments

                Comment on this article