      Single isocenter SRS using CAVMAT offers improved robustness to commissioning and treatment delivery uncertainty compared to VMAT

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          In this study, we evaluate and compare single isocenter multiple target VMAT (SIMT) and Conformal Arc Informed VMAT (CAVMAT) radiosurgery's sensitivity to uncertainties in dosimetric leaf gap (DLG) and treatment delivery. CAVMAT is a novel planning technique that uses multiple target conformal arcs as the starting point for limited inverse VMAT optimization.

          Methods

          All VMAT and CAVMAT plans were recalculated with DLG values of 0.4, 0.8, and 1.2 mm. DLG effect on V 6Gy[cc], V 12Gy[cc], and V 16Gy[cc], and target dose was evaluated. Plans were delivered to a Delta 4 (ScandiDos, Madison, WI) phantom and gamma analysis performed with varying criteria. Log file analysis was performed to evaluate MLC positional error. Sixteen targets were delivered to a SRS MapCHECK (Sun Nuclear Corp., Melbourne, FL) to evaluate VMAT and CAVMAT's dose difference (DD) as a function of DLG.

          Results

          VMAT's average maximum and minimum target dose sensitivity to DLG was 9.08 ±3.50%/mm and 9.50 ± 3.30%/mm, compared to 3.20 ± 1.60%/mm and 4.72 ± 1.60%/mm for CAVMAT. For VMAT, V 6Gy[cc], V 12Gy[cc], and V 16Gy[cc] sensitivity was 35.83 ± 9.50%/mm, 34.12 ± 6.60%/mm, and 39.23 ± 8.40%/mm. In comparison, CAVMAT's sensitivity was 23.19 ± 4.50%/mm, 22.45 ± 4.40%/mm, and 24.88 ± 4.90%/mm, respectively. Upon delivery to the Delta 4, CAVMAT offered superior dose agreement compared to VMAT. For a 1%/1 mm gamma analysis, VMAT and CAVMAT had a passing rate of 94.53 ± 4.40% and 99.28 ± 1.70%, respectively. CAVMAT was more robust to DLG variation, with the SRS MapCHECK plans yielding an absolute average DD sensitivity of 2.99 ± 1.30%/mm compared to 5.07 ± 1.10%/mm for VMAT. Log files demonstrated minimal differences in MLC positional error for both techniques.

          Conclusions

          CAVMAT remains robust to delivery uncertainties while offering a target dose sensitivity to DLG less than half that of VMAT, and 65% of that of VMAT for V 6Gy[cc], V 12Gy[cc], and V 16Gy[cc]. The superior dose agreement and reduced sensitivity of CAVMAT to DLG uncertainties indicate promise as a robust alternative to VMAT for SIMT SRS.

          Related collections

          Most cited references38

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Stereotactic radiosurgery for brain metastases: analysis of outcome and risk of brain radionecrosis

          Purpose to investigate the factors affecting survival and toxicity in patients treated with stereotactic radiosurgery (SRS), with special attention to volumes of brain receiving a specific dose (V10 - V16 Gy) as predictors for brain radionecrosis. Patients and Methods Two hundred six consecutive patients with 310 cerebral metastases less than 3.5 cm were treated with SRS as primary treatment and followed prospectively at University of Rome La Sapienza Sant'Andrea Hospital. Overall survival, brain control, and local control were estimated using the Kaplan-Meier method calculated from the time of SRS. Univariate and multivariate analysis using a Cox proportional hazards regression model were performed to determine the predictive value of prognostic factors for treatment outcome and SRS-related complications. Results Median overall survival and brain control were 14.1 months and 10 months, respectively. The 1-year and 2-year survival rates were 58% and 24%, and respective brain control were 43% and 22%. Sixteen patients recurred locally after SRS, with 1-year and 2-year local control rates of 92% and 84%, respectively. On multivariate analysis, stable extracranial disease and KPS >70 were associated with the most significant survival benefit. Neurological complications were recorded in 27 (13%) patients. Severe neurological complications (RTOG Grade 3 and 4) occurred in 5.8% of patients. Brain radionecrosis occurred in 24% of treated lesions, being symptomatic in 10% and asymptomatic in 14%. On multivariate analysis, V10 through V16 Gy were independent risk factors for radionecrosis, with V10 Gy and V12 Gy being the most predictive (p = 0.0001). For V10 Gy >12.6 cm3 and V12 Gy >10.9 cm3 the risk of radionecrosis was 47%. Conclusions SRS alone represents a feasible option as initial treatment for patients with brain metastases, however a significant subset of patients may develop neurological complications. Lesions with V12 Gy >8.5 cm3 carries a risk of radionecrosis >10% and should be considered for hypofractionated stereotactic radiotherapy especially when located in/near eloquent areas.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Feasibility of single-isocenter volumetric modulated arc radiosurgery for treatment of multiple brain metastases.

            To evaluate the relative plan quality of single-isocenter vs. multi-isocenter volumetric modulated arc therapy (VMAT) for radiosurgical treatment of multiple central nervous system metastases. VMAT plans were created using RapidArc technology for treatment of simulated patients with three brain metastases. The plans consisted of single-arc/single-isocenter, triple-arc (noncoplanar)/single-isocenter, and triple-arc (coplanar)/triple-isocenter configurations. All VMAT plans were normalized to deliver 100% of the 20-Gy prescription dose to all lesions. The plans were evaluated by calculation of Paddick and Radiation Therapy Oncology Group conformity index scores, Paddick gradient index scores, and 12-Gy isodose volumes. All plans were judged clinically acceptable, but differences were observed in the dosimetric parameters, with the use of multiple noncoplanar arcs showing small improvements in the conformity indexes compared with the single-arc/single-isocenter and triple-arc (coplanar)/triple-isocenter plans. Multiple arc plans (triple-arc [noncoplanar]/single-isocenter and triple-arc [coplanar]/triple-isocenter) showed smaller 12-Gy isodose volumes in scenarios involving three metastases spaced closely together, with only small differences noted among all plans involving lesions spaced further apart. Our initial results suggest that single-isocenter VMAT plans can be used to deliver conformity equivalent to that of multiple isocenter VMAT techniques. For targets that are closely spaced, multiple noncoplanar single-isocenter arcs might be required. VMAT radiosurgery for multiple targets using a single isocenter can be efficiently delivered, requiring less than one-half the beam time required for multiple isocenter set ups. VMAT radiosurgery will likely replace multi-isocenter techniques for linear accelerator-based treatment of multiple targets.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              A system for stereotactic radiosurgery with a linear accelerator.

              A small field irradiation technique to deliver high doses of single fraction photon radiation to small, precisely located volumes (0.5 to 8 cm3) within the brain has been developed. Our method uses a modified Brown-Roberts-Wells (BRS), CT-guided, stereotactic system and a 6 MV linear accelerator equipped with a special collimator (diameters of 12.5 mm to 30.0 mm projected to isocenter) located 23 cm from isocenter. Target localization via planar angiography has been added. Treatment consists of a series of arcing beams using both gantry and couch rotations. During treatment, the patient's head is immobilized independently of the radiotherapy couch and is precisely positioned without reference to room lasers or light field. A precise verification of alignment precedes each treatment. Extensive performance tests have shown that a target, localized by CT, can be irradiated with a positional accuracy of 2.4 mm in any direction with 95% confidence. If angiography is used for localization, the results are better. The dose 1.0 cm outside the target volume is less than 20% of the prescribed dose for a medium sized collimator.
                Bookmark

                Author and article information

                Contributors
                william.giles@duke.edu
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                24 June 2021
                July 2021
                : 22
                : 7 ( doiID: 10.1002/acm2.v22.7 )
                : 36-43
                Affiliations
                [ 1 ] Medical Physics Graduate Program Duke University Durham North Carolina USA
                [ 2 ] Department of Radiation Oncology Duke University Medical Center Durham North Carolina USA
                [ 3 ] Medical Physics Graduate Program Duke Kunshan University Suzhou China
                Author notes
                [*] [* ] Author to whom correspondence should be addressed. William Giles

                Email: william.giles@ 123456duke.edu

                Article
                ACM213248
                10.1002/acm2.13248
                8292691
                34165217
                87e8fa8f-97d5-4c65-961b-3209664cd77c
                © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine

                This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 12 January 2021
                : 18 September 2020
                : 15 March 2021
                Page count
                Figures: 5, Tables: 0, Pages: 8, Words: 5582
                Categories
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                July 2021
                Converter:WILEY_ML3GV2_TO_JATSPMC version:6.0.4 mode:remove_FC converted:21.07.2021

                cavmat,single isocenter,srs,vmat
                cavmat, single isocenter, srs, vmat

                Comments

                Comment on this article