1
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Technical evaluation of the cone‐beam computed tomography imaging performance of a novel, mobile, gantry‐based X‐ray system for brachytherapy

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          A novel, mobile cone‐beam computed tomography (CBCT) system for image‐guided adaptive brachytherapy was recently deployed at our hospital as worldwide first site. Prior to the device's clinical operation, a profound characterization of its imaging performance was conducted. This was essential to optimize both the imaging workflow and image quality for achieving the best possible clinical outcomes. We present the results of our investigations.

          Methods

          The novel CBCT‐system features a ring gantry with 121 cm clearance as well as a 43.2 × 43.2 cm 2 flat‐panel detector, and is controlled via a tablet‐personal computer (PC). For evaluating its imaging performance, the geometric reproducibility as well as imaging fidelity, computed tomography (CT)‐number accuracy, uniformity, contrast‐noise‐ratio (CNR), noise characteristics, and spatial resolution as fundamental image quality parameters were assessed. As dose metric the weighted cone‐beam dose index (CBDI w) was measured. Image quality was evaluated using standard quality assurance (QA) as well as anthropomorphic upper torso and breast phantoms. Both in‐house and manufacturer protocols for abdomen, pelvis, and breast imaging were examined.

          Results

          Using the in‐house protocols, the QA phantom scans showed altogether a high image quality, with high CT‐number accuracy ( R > 0.97) and uniformity (<12 Hounsfield Unit (HU) cupping), reasonable noise and imaging fidelity, and good CNR at bone–tissue transitions of up to 28:1. Spatial resolution was strongly limited by geometric instabilities of the device. The breast phantom scans fulfilled clinical requirements, whereas the abdomen and pelvis scans showed severe artifacts, particularly at air/bone–tissue transitions.

          Conclusion

          With the novel CBCT‐system, achieving a high image quality appears possible in principle. However, adaptations of the standard protocols, performance enhancements in image reconstruction referring to artifact reductions, as well as the extinction of geometric instabilities are imperative.

          Related collections

          Most cited references35

          • Record: found
          • Abstract: not found
          • Article: not found

          Practical cone-beam algorithm

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            What is cone-beam CT and how does it work?

            This article on x-ray cone-beam CT (CBCT) acquisition provides an overview of the fundamental principles of operation of this technology and the influence of geometric and software parameters on image quality and patient radiation dose. Advantages of the CBCT system and a summary of the uses and limitations of the images produced are discussed. All current generations of CBCT systems provide useful diagnostic images. Future enhancements most likely will be directed toward reducing scan time; providing multimodal imaging; improving image fidelity, including soft tissue contrast; and incorporating task-specific protocols to minimize patient dose.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Flat-panel cone-beam computed tomography for image-guided radiation therapy.

              Geometric uncertainties in the process of radiation planning and delivery constrain dose escalation and induce normal tissue complications. An imaging system has been developed to generate high-resolution, soft-tissue images of the patient at the time of treatment for the purpose of guiding therapy and reducing such uncertainties. The performance of the imaging system is evaluated and the application to image-guided radiation therapy is discussed. A kilovoltage imaging system capable of radiography, fluoroscopy, and cone-beam computed tomography (CT) has been integrated with a medical linear accelerator. Kilovoltage X-rays are generated by a conventional X-ray tube mounted on a retractable arm at 90 degrees to the treatment source. A 41 x 41 cm(2) flat-panel X-ray detector is mounted opposite the kV tube. The entire imaging system operates under computer control, with a single application providing calibration, image acquisition, processing, and cone-beam CT reconstruction. Cone-beam CT imaging involves acquiring multiple kV radiographs as the gantry rotates through 360 degrees of rotation. A filtered back-projection algorithm is employed to reconstruct the volumetric images. Geometric nonidealities in the rotation of the gantry system are measured and corrected during reconstruction. Qualitative evaluation of imaging performance is performed using an anthropomorphic head phantom and a coronal contrast phantom. The influence of geometric nonidealities is examined. Images of the head phantom were acquired and illustrate the submillimeter spatial resolution that is achieved with the cone-beam approach. High-resolution sagittal and coronal views demonstrate nearly isotropic spatial resolution. Flex corrections on the order of 0.2 cm were required to compensate gravity-induced flex in the support arms of the source and detector, as well as slight axial movements of the entire gantry structure. Images reconstructed without flex correction suffered from loss of detail, misregistration, and streak artifacts. Reconstructions of the contrast phantom demonstrate the soft-tissue imaging capability of the system. A contrast of 47 Hounsfield units was easily detected in a 0.1-cm-thick reconstruction for an imaging exposure of 1.2 R (in-air, in absence of phantom). The comparison with a conventional CT scan of the phantom further demonstrates the spatial resolution advantages of the cone-beam CT approach. A kV cone-beam CT imaging system based on a large-area, flat-panel detector has been successfully adapted to a medical linear accelerator. The system is capable of producing images of soft tissue with excellent spatial resolution at acceptable imaging doses. Integration of this technology with the medical accelerator will result in an ideal platform for high-precision, image-guided radiation therapy.
                Bookmark

                Author and article information

                Contributors
                andre.karius@uk-erlangen.de
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                14 December 2021
                February 2022
                : 23
                : 2 ( doiID: 10.1002/acm2.v23.2 )
                : e13501
                Affiliations
                [ 1 ] Department of Radiation Oncology Universitätsklinikum Erlangen Friedrich‐Alexander Universität Erlangen‐Nürnberg Universitätsstraße 27 Erlangen Germany
                [ 2 ] Comprehensive Cancer Center Erlangen‐EMN (CCC ER‐EMN) Erlangen Germany
                [ 3 ] Institute of Medical Physics Friedrich‐Alexander‐University Erlangen‐Nuremberg Henkestraße 91 Erlangen Germany
                Author notes
                [*] [* ] Correspondence

                Andre Karius, Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich‐Alexander Universität Erlangen‐Nürnberg, Universitätsstraße 27, 91054 Erlangen, Germany.

                Email: andre.karius@ 123456uk-erlangen.de

                Article
                ACM213501
                10.1002/acm2.13501
                8833290
                34905285
                932bd286-5539-4724-9b92-14db9301b63b
                © 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine

                This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 21 October 2021
                : 31 July 2021
                : 29 November 2021
                Page count
                Figures: 9, Tables: 3, Pages: 17, Words: 11028
                Categories
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                February 2022
                Converter:WILEY_ML3GV2_TO_JATSPMC version:6.1.1 mode:remove_FC converted:11.02.2022

                cone‐beam computed tomography (cbct),image‐guided adaptive brachytherapy,image quality,imaging performance analysis,mobile x‐ray system

                Comments

                Comment on this article