      Evaluation of Surface Changes of Stainless Steel Miniplates and Screws Following Retrieval from Maxillofacial Trauma and Orthognathic Surgery Patients: A Comparative Study

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          A BSTRACT

          Background:

          This study examines the surface changes of stainless steel miniplates and screws after their retrieval from patients who underwent maxillofacial trauma and orthognathic surgery. The assessment focuses on comparing the alterations in these materials, aiming to contribute to our understanding of their durability and performance in clinical settings

          Materials and Methods:

          A total of 60 stainless steel miniplates and screws were collected from 30 patients who had previously undergone either maxillofacial trauma or orthognathic surgery. The retrieved miniplates and screws were carefully removed from the patients and cleaned to remove any organic debris. Each specimen was then examined for surface changes. Surface changes were evaluated using visual inspection, scanning electron microscopy (SEM), and roughness measurements. Visual inspection provided a qualitative assessment, while SEM allowed for a more detailed examination of the surfaces. Roughness measurements were conducted using a profilometer.

          Results:

          Visual inspection revealed varying degrees of surface changes in the retrieved miniplates and screws. These changes included scratches, corrosion, and discoloration. SEM analysis confirmed the presence of surface alterations, with some specimens showing more significant damage, such as pitting and cracks. Roughness measurements indicated an increase in surface roughness for both miniplates and screws, suggesting that the surfaces had become less smooth

          Conclusion:

          This comparative study of stainless steel miniplates and screws retrieved from maxillofacial trauma and orthognathic surgery patients demonstrated that these materials undergo surface changes over time.

          Related collections

          Most cited references5

          • Record: found
          • Abstract: found
          • Article: not found

          A comparison of biocompatibility and osseointegration of ceramic and titanium implants: an in vivo and in vitro study.

          This study compared the biocompatibility in vitro and the osseointegration in vivo of zirconium and titanium implants regarding implant surfaces and the bone-implant contacts. The different implant surfaces and the biocompatibility of zirconium versus titanium implants were determined by vitality and cytotoxic tests in vitro. The contact of the osteoblasts to the implant surface was determined by scanning electron microscopy (SEM). The in vivo study for osseointegration was performed in domestic pigs over 4 and 12 weeks. In each animal, 4 zirconium and 4 titanium implants (WhiteSky, BlueSky, Bredent, Germany) were inserted in the os frontale and analysed by histomorphometry. Cytotoxicity and SEM showed good biocompatibility in relation to the investigated implant materials. Histological results showed direct bone-implant contact of the implant surfaces. The zirconium implants showed a slight delay in osseointegration in terms of bone-implant contact as measured by histomorphometry (after 4 weeks, zirconium (59.3 ± 4.6%) versus titanium (64.1 ± 3.9%); after 12 weeks, zirconium (67.1 ± 2.3%) versus titanium (73.6 ± 3.2%). A statistically significant difference between the two groups was not observed. The results indicated similar biocompatibility and osseointegration for zirconium compared to titanium implants. Copyright © 2012 International Association of Oral and Maxillofacial Surgeons. Published by Elsevier Ltd. All rights reserved.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Rigid fixation of mandibular condyle fractures.

            This article reviews the anatomy and surgical approaches for treating fractures of the mandibular condyle with plate and screw fixation. Advantages and disadvantages of the preauricular, submandibular, intraoral, retromandibular, and rhytidectomy approaches are presented.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Mechanical properties of orthodontic wires in tension, bending, and torsion.

              The mechanical properties of three sizes of stainless steel (SS), nickel-titanium (NT), and titanium-molybdenum (TM) orthodontic wires were studied in tension, bending, and torsion. The wires (0.016 inch, 0.017 by 0.025 inch, and 0.019 by 0.025 inch) were tested in the as-received condition. Tensile testing and stiffness testing machines along with a torsional instrument were used. Mean values and standard deviations of properties were computed. The data were analyzed statistically by analysis of variance using a factorial design. Means were ranked by a Tukey interval calculated at the 95 percent level of confidence. In tension, the stainless steel wires had the least maximum elastic strain or springback, whereas the titanium-molybdenum wires had the most. Higher values of springback indicate the capacity for an increased range of activation clinically. In bending and torsion, the stainless steel wires had the least stored energy at a fixed moment, whereas the nickel-titanium wires had the most. Spring rates in bending and torsion, however, were highest for stainless steel wires and lowest for nickel-titanium wires. A titanium-molybdenum teardrop closing loop delivered less than one half the force of a comparable stainless steel loop for similar activations.
                Bookmark

                Author and article information

                Journal
                J Pharm Bioallied Sci
                J Pharm Bioallied Sci
                JPBS
                J Pharm Bioall Sci
                Journal of Pharmacy & Bioallied Sciences
                Wolters Kluwer - Medknow (India )
                0976-4879
                0975-7406
                February 2024
                20 December 2023
                : 16
                : Suppl 1
                : S835-S837
                Affiliations
                [1 ]Department of Oral and Maxillofacial Surgery, Institute of Dental Sciences, Bareilly, Uttar Pradesh, India
                [2 ]Department of Periodontology, D J College of Dental Sciences and Research, Modinagar, Uttar Pradesh, India
                [3 ]Department of Oral and Maxillofacial Surgery, Ambikapur, Chhattisgarh, India
                [4 ]Department of Oral and Maxillofacial Surgery, HT Hospital Cosmetic and Trauma Centre, Bhopal, Madhya Pradesh, India
                [5 ]Department of Oral and Maxillofacial Surgery, Bharati Vidyapeeth (Deemed to be University), Dental College and Hospital, Navi Mumbai, Maharashtra, India
                [6 ]Department of Orthodontics, Bharati Vidyapeeth (Deemed to be University), Dental College and Hospital, Navi Mumbai, Maharashtra, India
                [7 ]Department of Oral Medicine and Radiology, New Horizon Dental College and Research Institute, Sakri Bilaspur, Chhattisgarh, India
                Author notes
                Address for correspondence: Dr. Himanshu P. Singh, Department of Oral and Maxillofacial Surgery, Institute of Dental Sciences, Bareilly - 243 006, Uttar Pradesh, India. E-mail: drhimanshupratap@ 123456gmail.com
                Article
                JPBS-16-835
                10.4103/jpbs.jpbs_1049_23
                11000920
                38595370
                9abc0872-0c6a-449b-b385-37333f98fd67
                Copyright: © 2023 Journal of Pharmacy and Bioallied Sciences

                This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.

                History
                : 14 October 2023
                : 19 October 2023
                : 24 October 2023
                Categories
                Original Article

                Pharmacology & Pharmaceutical medicine
                clinical practice,durability,maxillofacial trauma,miniplates,orthognathic surgery,retrieval,screws,stainless steel,surface analysis,surface changes