      Microwave ablation with a single small-gauge triaxial antenna: in vivo porcine liver model.

      Radiology
      Animals, Electrocoagulation, instrumentation, methods, Equipment Design, Female, Liver, pathology, surgery, Liver Circulation, physiology, Microwaves, therapeutic use, Models, Animal, Swine, Time Factors

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To evaluate the performance of a 17-gauge triaxial antenna at microwave ablation in an in vivo porcine liver model. This study was approved by the institutional animal care and use committee. Thirteen female domestic pigs (mean weight, 45 kg) were used. Ablations were performed with a prototype microwave ablation system and triaxial antenna by using a constant, continuous-wave power of 68 W for 2 (n = 6), 3 (n = 6), 4 (n = 6), 5 (n = 6), 6 (n = 6), 7 (n = 13), 10 (n = 7), and 12 (n = 8) minutes. Animals were euthanized after ablation, livers were removed, and ablation zones were sliced and measured for size and roundness. A mixed linear model with animals modeled as random effects was used to test for significant differences in ablation zone metrics among time groups; post hoc tests were used to detect significant differences between time groups. Mean ablation zone diameters ranged from 2.05 cm +/- 0.23 (standard deviation) at 2 minutes to 2.59 cm +/- 0.53 at 12 minutes. Thirteen (32%) of 40 ablation zones with mean maximum diameters greater than 3.0 cm were observed at the 5-12-minute time groups. No significant differences in ablation zone diameter were observed among all groups (P > .05), but a trend of increasing diameter with time was noted. Mean isoperimetric ratio (a measure of roundness) for all ablation zones was 0.88 +/- 0.02, which indicates minimal heat sinking near vessels. The triaxial microwave ablation system is capable of creating relatively large, circular zones of ablation in minutes with minimal effects from local blood flow. (c) RSNA, 2007.

          Related collections

          Author and article information

          Journal
          17255414
          1945245
          10.1148/radiol.2422051411

          Chemistry
          Animals,Electrocoagulation,instrumentation,methods,Equipment Design,Female,Liver,pathology,surgery,Liver Circulation,physiology,Microwaves,therapeutic use,Models, Animal,Swine,Time Factors

          Comments

          Comment on this article