The perioperative management of dabigatran is challenging, and recommendations based on activated partial thromboplastin time (aPTT) and thrombin time (TT) are unsatisfactory. Dedicated coagulation tests have limitations at plasma concentrations < 50 ng/ml. Therefore, a more sensitive test, which is available 24/7, is required. It was the aim of this study to investigate the performance of the Hemoclot Thrombin Inhibitors® LOW (HTI LOW) kit, a diluted thrombin time, and the STA® - ECA II(ECA-II) kit, a chromogenic variant of the ecarin clotting time, that were developed to measure low dabigatran concentrations, compared to reference dabigatran analysis by liquid chromatography tandem mass-spectrometry (LC-MS/MS). This study included 33 plasma samples from patients treated with dabigatran etexilate who had plasma concentrations < 200 ng/ml. HTI LOW and ECA-II were performed along with HTI, aPTT (STA®-C. K.Prest® and SynthasIL®) and TT (STA® - Thrombin). All procedures were performed according to recommendations by the manufacturers. Linear (or curvilinear) correlations and Bland-Altman analyses were calculated. For free dabigatran concentrations < 50 ng/ml, the R² of linear correlations were 0.69, 0.84 and 0.61, with HTI, HTI LOW and ECA-II, respectively. The R² for TT, STA®-C. K.Prest® and SynthasIL® were 0.67, 0.42 and 0.15. For HTI, HTI LOW and ECA-II, Bland-Altman analyses revealed mean differences of -6 ng/ml (95 %CI: -25-14 ng/ml), 1 ng/ml (95 %CI: -18-19 ng/ml) and -1 ng/ml (95 %CI: -25-23 ng/ml), demonstrating that tests dedicated to measuring low concentrations are more accurate than HTI. In conclusion, the use of HTI LOW or ECA-II to assess low plasma dabigatran concentrations is supported by our findings.