      Color and conventional image-directed Doppler ultrasonography: accuracy and sources of error in quantitative blood flow measurements.

      Journal of clinical ultrasound : JCU
      Blood Flow Velocity, Humans, Models, Anatomic, Pulsatile Flow, Reproducibility of Results, Ultrasonography, methods

      Read this article at

      ScienceOpenPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Accuracy of two systems--conventional (DRF 400, Diasonics) and color-coded (Angiodynograph, Quantum/Phillips) image-directed Doppler ultrasonography--was investigated using an in vitro model that generated both monophasic and triphasic pulsatile flow patterns. Estimated and actual blood volume flow rates showed good correlations, but the sampling with a hand-held transducer led to wide variations in measurement error for the conventional (-69.2% to 50%) and the color-coded (-79.3% to 265.7%) systems. By performing multiple measurements, one could improve accuracy considering only the maximal values of a series instead of the mean values. Accuracy was impaired by interposed muscular or fatty tissue due to false low time-average velocity measurements caused by a loss of Doppler signal. Comparison of both systems revealed significant differences between pulsatility index values (p less than 0.001), blood flow velocities (p less than 0.001), and blood volume flow rates (p less than 0.05 for program flow, p less than 0.001 for manual and automatic flow program of the color-coded system).

          Related collections

          Author and article information

          Comments

          Comment on this article