      Multigap RPC for PET: development and optimisation of the detector design

      Preprint

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Transforming the resistive plate chambers from charged-particle into gamma-quanta detectors opens the way towards their application as a basic element of a hybrid imaging system, which combines positron emission tomography (PET) with magnetic resonance imaging (MRI) in a single device and provides non- and minimally- invasive quantitative methods for diagnostics. To this end, we performed detailed investigations encompassing the whole chain from the annihilation of the positron in the body, through the conversion of the created photons into electrons and to the optimization of the electron yield in the gas. GEANT4 based simulations of the efficiency of the RPC photon detectors with different converter materials and geometry were conducted for optimization of the detector design. The results justify the selection of a sandwich-type gas-insulator-converter design, with Bi or Pb as converter materials.

          Related collections

          Most cited references7

          • Record: found
          • Abstract: found
          • Article: not found

          PET/MRI: the blended-modality choice of the future?

          This article addresses the emerging technology of PET coupled with MRI, or PET/MRI, which could become the technology of choice in the future for many reasons. Some of these reasons will be discussed, along with a historical account of the field of MRI and how this modality has evolved to include many aspects of molecular and functional imaging. After reading this article, nuclear medicine technologists should be able to provide an overview of the history of MRI, discuss PET and how it is mainly used today melded to CT as PET/CT, discuss how MRI is used diagnostically, explain how PET technology and MRI technology are able to function simultaneously together as PET/MRI, discuss some issues concerning who will operate these new units, and discuss the possibility that PET/MRI could be the blended technology of choice in the future.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            PET energy-based scatter estimation and image reconstruction with energy-dependent corrections.

            In this paper we propose a comprehensive energy-based scatter correction approach for positron emission tomography (PET). We take advantage of the marked difference between the energy spectra of the unscattered and scattered photons, and use the detailed energy information that comes with the list-mode data for the estimation of the scattered events distribution in the data space. Also, inside the maximum-likelihood expectation maximization (ML-EM) image reconstruction algorithm, we introduce energy-dependent factors that individualize the correction terms for each event, given its position and energy information. The central piece of our approach is the two-dimensional detector energy response model represented as a linear combination of four components, each one representing a particular state a PET event can be found in: both photons unscattered, the second scattered while the first not, the first photon scattered while the second not and both photons scattered. For a set of events collected in the vicinity of a point in the projection space, the coefficient of each component is determined by applying a statistical estimator. As a result we obtain the number of scattered events that are in the given set. The model also gives us the variation of scatter fraction with the photon pair energies for that particular position in the data space. A simulation study that demonstrates the proposed methods is presented.
              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              A 20ps timing device—A Multigap Resistive Plate Chamber with 24 gas gaps

                Bookmark

                Author and article information

                Journal
                21 December 2012
                Article
                10.1088/1748-0221/8/11/P11011
                1212.5551
                68d72358-5146-43b0-92f5-f6ea5abf9880

                http://arxiv.org/licenses/nonexclusive-distrib/1.0/

                History
                Custom metadata
                to be published in JINST
                physics.ins-det physics.med-ph

                Comments

                Comment on this article