          Background: Early diagnosis of ischemia is complicated by the poor sensitivity of standard tests and contraindication for stress testing in unstable angina patients. Magnetocardiography (MCG) imaging can be used for the rapid, non-invasive detection of ischemia at rest. Methods: We studied 125 patients with presumed ischemic chest pain. All were chest pain free at the time of scanning. A 6-minute resting MCG scan (CardioMag Imaging, Inc., New York, 9-channel system) was performed. Following the MCG scan, automated software data analysis was performed, and quantitative scores were automatically calculated for each subject. The presence of ischemia was determined after testing with serial troponins, stress testing, and/or coronary angiography. Results: The mean age was 59.4 ± 13.6 years. Most patients (86.4%) had non-ischemic 12-lead ECG and normal troponin (86.2%). Fifty-five patients (44.0%) were determined to be ischemic. The MCG sensitivity, specificity, positive and negative predictive value was 76.4, 74.3, 70.0 and 80.0%, respectively, for the detection of ischemia (p < 0.0001). Conclusions: MCG is a new rapid, non-invasive imaging tool able to detect repolarization abnormalities at rest consistent with ischemia in patients presenting with chest pain syndrome and normal or non-specific 12-lead ECG and normal troponin.

          Related collections

          Most cited references 7

          • Record: found
          • Abstract: not found
          • Article: not found

          MAGNETOCARDIOGRAMS TAKEN INSIDE A SHIELDED ROOM WITH A SUPERCONDUCTING POINT‐CONTACT MAGNETOMETER

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Comparative capabilities of electrocardiography and magnetocardiography

              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              Magnetocardiographic localization of arrhythmia substrates: a methodology study with accessory pathway ablation as reference

                Bookmark

                Author and article information

                Journal
                CRD
                Cardiology
                10.1159/issn.0008-6312
                Cardiology
                S. Karger AG
                0008-6312
                1421-9751
                2006
                November 2006
                15 November 2006
                : 106
                : 4
                : 270-276
                Affiliations
                aCedars-Sinai Medical Center, Division of Cardiology, Los Angeles, Calif., and bMayo Clinic College of Medicine, Department of Emergency Medicine, Rochester, Minn., USA; cKlinikum Hoyerswerda, Division of Cardiology, Hoyerswerda, Germany
                Article
                93490 Cardiology 2006;106:270–276
                10.1159/000093490
                16733351
                © 2006 S. Karger AG, Basel

                Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher. Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in government regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug. Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.

                Page count
                Figures: 3, Tables: 3, References: 27, Pages: 7
                Categories
                Original Research

                Comments

                Comment on this article