      Can 'extrafine' dry powder aerosols improve lung deposition?

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          There is increasing interest in the use of so-called 'extrafine' aerosols to target the small airways in the management of asthma and COPD. Using previously presented deposition data, we assessed whether submicron (<1μm) particles can improve central and deep lung deposition. Our data show instead that particles in the range 1-3μm are much more relevant in this respect. Based on this finding the Symbicort Turbuhaler, Seretide Diskus, Rolenium Elpenhaler and Foster (Fostair) NEXThaler ICS/LABA combination DPIs were tested in vitro as a function of the pressure drop (2, 4 and 6kPa) across the inhaler. Obtained fine particle fractions (FPFs) <5μm (as percent of label claim) were divided into subfractions <1, 1-3 and 3-5μm. Differences of up to a factor of 4 were found between the best (Turbuhaler) and worst performing DPI (Elpenhaler), particularly for the FPF in the size range 1-3μm. The NEXThaler, described as delivering 'extrafine' particles, did not appear to be superior in this size range. The marked differences in amount and size distribution of the aerosols between the devices in this study must cause significant differences in the total lung dose and drug distribution over the airways.

          Related collections

          Author and article information

          Journal
          Eur J Pharm Biopharm
          European journal of pharmaceutics and biopharmaceutics : official journal of Arbeitsgemeinschaft fur Pharmazeutische Verfahrenstechnik e.V
          Elsevier BV
          1873-3441
          0939-6411
          Oct 2015
          : 96
          Affiliations
          [1 ] Department of Pharmaceutical Technology and Biopharmacy, University of Groningen, Ant. Deusinglaan 1, Groningen, The Netherlands. Electronic address: a.h.de.boer@rug.nl.
          [2 ] Department of Pharmaceutical Technology and Biopharmacy, University of Groningen, Ant. Deusinglaan 1, Groningen, The Netherlands.
          Article
          S0939-6411(15)00316-1
          10.1016/j.ejpb.2015.07.016
          26220014
          90258f29-cee6-4720-adf4-620bb8dd59d8
          History

          Lung deposition,Small airway targeting,Submicron particles,Dry powder inhaler,Extrafine aerosol,ICS

          Comments

          Comment on this article