Introduction
In September 2012, the United Stated Food and Drug Administration approved the use
of a fully subcutaneous implantable cardiac defibrillator (S-ICD, Boston Scientific
Inc). The device is implanted in the left midaxillary space and attached to a single
lead that is tunneled subcutaneously from the xyphoid process in 2 directions, superiorly
to the sternal manubrium joint to the left of the sternum and laterally to the pulse
generator. The lead consists of a single coil in the portion of lead along the sternum
and 2 sensing electrodes, 1 at the tip of the lead at the upper portion of the sternum
and 1 at the xyphoid process. Sensing is achieved via 1 of 3 potential configurations:
between the device and the lower electrode, between the device and the upper electrode,
or between the 3 electrodes. The sensing vector is automatically chosen by the device
to minimize the chance of T-wave oversensing, but it can be manually overriden.
1
A deep brain stimulator (DBS) is an electronic device consisting of a pulse generator
and 1 or more electrodes implanted in the brain. It can be programmed to operate in
a bipolar or unipolar stimulation mode. It is used for the treatment of Parkinson
disease, among other neurologic conditions.
2
Manufacturer’s recommendations for concomitant use of a transvenous implantable cardiac
defibrillator (ICD) and DBS include setting the ICD to bipolar sensing. Sensing in
an S-ICD is achieved via much wider bipoles than in a transvenous ICD, raising the
concern of adverse interaction between the 2 devices. To our knowledge, we present
the first case report of successful implantation of an S-ICD in a patient with a previously
implanted DBS.
KEY TEACHING POINTS
•
The subcutaneous cardiac defibrillator (S-ICD) represents a major advance in ICD technology
with the ability to provide sudden death prevention without transvenous leads. Because
of its wide sensing bipole, interaction with other implanted electronic devices is
a concern. This includes patients with a deep brain stimulator (DBS), which is used
for treatment of neurologic disorders such as Parkinson disease.
•
Implantation of an S-ICD in patients with a preexisting DBS requires a multidisciplinary
approach with the patient’s neurologist for programming the DBS to a bipolar mode
if possible to limit the possibility of interaction with the S-ICD. In addition, technical
support should be available during S-ICD implantation to test sensing with different
DBS settings and for interrogation of the DBS after defibrillation threshold testing.
•
This case report outlines an approach that was successful when both devices coexisted
in the same patient without any adverse effect on the S-ICD or the patient’s neurologic
symptoms.
Case report
A 51-year-old man presented as an outpatient to our institution for consideration
of an S-ICD implantation. His past medical history consisted of coronary artery disease,
for which he had undergone placement of multiple coronary stents, and early-onset
Parkinson disease, for which he had undergone implantation of a Medtronic Activa DBS
in the right prepectoral area. In 1996, he had an episode of polymorphic ventricular
tachycardia, which resulted in cardiac arrest. At that time, a single-chamber ICD
was implanted in the left prepectoral area for secondary prevention of sudden cardiac
death. His left ventricular ejection fraction was and remains normal. Between 1996
and 2005, he underwent 4 ICD generator replacements. His initial right ventricular
lead was a Ventritex Cadence single-coil lead, which failed and was replaced by a
dual-coil St. Jude Medical Riata lead in 2005. In view of the recent Food and Drug
Administration recommendation,
3
the patient underwent routine surveillance imaging of the lead at another institution
and was found to have externalization of a conductor on fluoroscopy. In addition,
an acute rise in the right ventricular threshold from 1 to 3.5 V was noted.
Because of a lack of confidence in the reliability of the Riata lead and the patient’s
desire to continue to have protection from sudden cardiac death, the patient was given
multiple options, including extraction of the transvenous lead and implantation of
new transvenous lead, or abandonment of the leads and implantation of an S-ICD. The
decision-making was complicated by the presence of the Medtronic Activa DBS, which
had provided him with significant relief from parkinsonian symptoms.
In patients with Parkinson disease, the DBS works by bilateral stimulation of the
internal globus pallidus or the subthalamic nucleus. Our patient had a single unit
with 2 leads, 1 to each cerebral hemisphere. Each lead has 4 electrodes, and the device
can be programmed to stimulate in either a unipolar or bipolar fashion. The device
can be programmed to a voltage mode or a current mode, and it can deliver 2 to 250
Hz at a pulse width of 60 to 450 μs and up 10.5 V (voltage mode) or 25.5 mA (current
mode).
The patient’s DBS had been chronically programmed to unipolar stimulation between
the DBS pulse generator and the lead(s). As a first step to facilitate S-ICD implantation,
we requested that the DBS be changed to a bipolar mode. Symptom relief from parkinsonian
symptoms persisted in bipolar mode.
The 2 DBS leads in the patient’s DBS can be programmed independently. The left hemisphere
lead was programmed to an output of 3.V, and the right hemisphere lead was programmed
to 2.1 V. The pulse width and frequency of both leads were the same at 90 μs and 180
Hz, respectively. During implantation of the S-ICD and defibrillation threshold (DFT)
testing, these settings were not manipulated.
Avoidance of T-wave oversensing by an S-ICD requires screening surface ECG recordings
simulating the sensing vectors of the S-ICD. Application of a template provided by
the manufacturer determines eligibility, which was adequate in this patient.
The patient was taken to the electrophysiology laboratory for implantation of the
S-ICD. A programmer and a technician were available to alter the programming of the
DBS as needed. The procedure was performed with the patient under general anesthesia.
The S-ICD implantation technique has been described elsewhere.
4
We performed the standard technique with a modification: we used a sheath in conjunction
with the tunneling tool to place the lead along the left side of the sternum, which
avoids the superior third incision.
After implantation, we tested for interaction of the S-ICD and the DBS. Changing between
unipolar and bipolar stimulation on the DBS was immediately apparent on the surface
ECG (Figure 1).
The S-ICD sensing vectors were recorded with the DBS in both unipolar and bipolar
configurations. There was no oversensing of DBS activity (in both bipolar and unipolar
modes) by the S-ICD (Figure 2). DFT testing was performed with successful sensing
and termination of induced ventricular fibrillation at 65, 50, and 35 J. The DBS was
active in bipolar mode during DFT testing. We interrogated the DBS after DFT testing
and found no interruption in normal function.
On routine follow-up 12 months after implantation, the patient was doing well with
no complications. He had not received any S-ICD shocks. Chest x-ray film showed a
well-positioned S-ICD device and electrode (Figure 3).
Discussion
DBS is an increasingly common treatment for a variety of neurologic disorders, including
Parkinson disease, so the possibility of a patient requiring both an ICD and a DBS
is increasing. Three previous case reports have documented the safety and lack of
interaction between transvenous ICDs and DBS. However, 1 case report did document
resetting of a DBS to an off mode after DFT testing of a transvenous ICD.
5
To our knowledge, this is the first report of the safe implantation and follow-up
of an S-ICD with a DBS. There was no issue acutely with interaction of the DBS and
ICD with the DBS in either bipolar or unipolar mode, and in all 3 sensing vectors
of the S-ICD. The lack of DBS artifact on the S-ICD (even with unipolar DBS stimulation)
likely is due to signal filtering in the S-ICD. The DBS was programmed to a stimulation
frequency of 180 Hz both chronically and during the implantation. The S-ICD allows
frequencies between only 3 and 40 Hz to pass and thus eliminates the DBS signal. In
contrast, the recording system in the electrophysiology laboratory where the case
was performed allows frequencies of 30 to 250 Hz to pass, making DBS unipolar stimulation
apparent on the surface ECG. Bipolar DBS stimulation was not seen on the ECG recording,
probably because it was of much lower amplitude than unipolar stimulation. Sensing
of ventricular fibrillation by the S-ICD was unaffected by active bipolar stimulation
from the DBS. In addition, the 3 ICD shocks delivered for DFT testing did not adversely
affect the DBS.
It is important to note that, during follow-up, the DBS was left in bipolar mode and
the S-ICD remained in its automatically selected ideal sensing vector. It is not clear
from this report whether with the DBS in unipolar mode or a different S-ICD sensing
vector and the DBS in either mode that oversensing and interaction may not have occurred
during follow-up. DBS devices can also be programmed to frequencies that are well
within the filter pass range of the S-ICD, which may result in a higher risk of interaction.
In addition, it is not clear whether S-ICD sensing chronically or DBS function after
an S-ICD shock would be unaffected if the DBS were on the left side of the patient.
An additional issue that may arise with the combination of these 2 devices is the
manner in which the DBS behaves after a power on reset event, which theoretically
can occur after an ICD shock. The current generation of Medtronic devices, 1 of which
was present in this patient, resets to the previously programmed parameters, even
after multiple resets. However, older generations of DBS devices (which still are
available but infrequently used) will revert to the default settings, which vary,
but do include, in some instances, a stimulation frequency of 30 Hz, which is within
the pass filters limits of the S-ICD and would increase the risk of oversensing of
DBS stimulation by the ICD. For patients with an S-ICD, such older-generation DBS
devices probably should be avoided. For patients with preexisting older-generation
DBS, this issue should be taken into account when considering a new S-ICD implantation.
Given the inherent limitations of a single case report, caution should be exercised
in applying these findings to similar clinical situations. Ideally, a series of such
cases would be useful to better understand the potential interactions and issues that
may arise when these devices coexist in a single patient.